All-Polymer Ultrasonic Transducer Design for an Intravascular Ultrasonography Application

Loading...
Publication Logo

Date

2019

Journal Title

Journal ISSN

Volume Title

Publisher

Tubitak Scientific & Technological Research Council Turkey

Open Access Color

GOLD

Green Open Access

Yes

OpenAIRE Downloads

171

OpenAIRE Views

166

Publicly Funded

No
Impulse
Average
Influence
Average
Popularity
Average

Research Projects

Journal Issue

Abstract

Intravascular ultrasonography (IVUS), a medical imaging modality, is used to obtain cross-sectional views of blood vessels from inside. In IVUS, transducers are brought to the proximity of the imaging targets so that high-resolution images can be obtained at high frequency without much concern of signal attenuation. To eliminate mechanical rotation rendered in conventional IVUS, it is proposed to manufacture a transducer array on a flexible substrate and wrap it around a cylindrical frame. The transducer of consideration is a capacitive micromachined ultrasonic transducer (CMUT). The whole device needs to be made out of polymers to be able to endure a high degree of bending (radius: 1 mm) Bending of the devices leads to considerable changes in the device characteristics, including resonant frequency and pull-in voltage due to geometrical dimension changes and stress induced. The main purpose of this work is to understand the effect of bending on the device characteristics by means of finite element analysis. Another objective of the work is to understand the relationships between such an effect and the device geometries. It is learned that the bending-induced stress depends strongly on anchor width, membrane thickness, and substrate thickness. It is also learned that resonant frequency and pull-in voltage become lower in most cases because of using a flexible substrate in comparison to those of the device on a rigid substrate. Bending-induced stress increases the spring constant and hence increases resonant frequency and pull-in voltage, although this effect is relatively weaker. For most of the device geometries, pull-in voltage is too high for the polymer material to endure. This is the main drawback of the all-polymer CMUT. In order to meet the design goal of 20 MHz resonant frequency, the membrane radius has to be smaller than 7.7 mu m for a thickness of 3 mu m.

Description

Hah, Dooyoung/0000-0002-1290-0597

Keywords

Polymer Microelectromechanical Systems, Ultrasonic Transducer, Flexible Substrate, Intravascular Ultrasonography, Polymer microelectromechanical systems, ultrasonic transducer, flexible substrate, intravascular ultrasonography

Fields of Science

0103 physical sciences, 01 natural sciences

Citation

WoS Q

Q3

Scopus Q

Q2
OpenCitations Logo
OpenCitations Citation Count
4

Source

Turkish Journal of Electrical Engineering and Computer Sciences

Volume

27

Issue

4

Start Page

2444

End Page

2455
PlumX Metrics
Citations

CrossRef : 3

Scopus : 4

Captures

Mendeley Readers : 11

Google Scholar Logo
Google Scholar™
OpenAlex Logo
OpenAlex FWCI
0.508

Sustainable Development Goals