Design of capacitive micromachined ultrasonic transducers (CMUTs) on a flexible substrate for intravascular ultrasonography (IVUS) applications

dc.contributor.author Hah, Dooyoung
dc.contributor.author Je, Chang Han
dc.contributor.author Lee, Sung-Q
dc.contributor.authorID 0000-0002-1290-0597 en_US
dc.contributor.department AGÜ, Mühendislik Fakültesi, Elektrik - Elektronik Mühendisliği Bölümü en_US
dc.date.accessioned 2021-07-28T10:36:33Z
dc.date.available 2021-07-28T10:36:33Z
dc.date.issued 2017 en_US
dc.description This work was partially supported by Research Fund of the Abdullah Gul University (Project Number: FOA-2016-49). en_US
dc.description.abstract Effects of substrate bending to the characteristics of capacitive miniaturized ultrasonic transducers (CMUTs) on a flexible substrate are studied through finite element analysis (FEA) for the design purpose. The target application of the devices is intravascular ultrasonography (IVUS) where transducers are brought to the proximity of the imaging targets so that high resolution images can be obtained without much concern of signal attenuation. In order to eliminate mechanical rotation used in the conventional IVUS, the transducer array can be manufactured on a flexible substrate and to wrap it around a cylindrical frame. It can be anticipated that the characteristics of the transducers will be altered by such bending of the substrate through geometrical dimension changes and stress induced. Pullin voltages and resonant frequencies of CMUTs were studied via FEA for various bending radii and membrane thicknesses. It was found that both pull-in voltages and resonant frequencies become smaller for the transducers on a bent substrate compared to the ones on a flat substrate. It was also found that pull-in voltages decrease as the substrate bending radius is reduced. en_US
dc.description.sponsorship Abdullah Gul University FOA-2016-49 en_US
dc.identifier.isbn 978-1-5386-2952-9
dc.identifier.uri https://hdl.handle.net/20.500.12573/879
dc.language.iso eng en_US
dc.publisher IEEE345 E 47TH ST, NEW YORK, NY 10017 USA en_US
dc.relation.journal 2017 SYMPOSIUM ON DESIGN, TEST, INTEGRATION AND PACKAGING OF MEMS/MOEMS (DTIP 2017) en_US
dc.relation.publicationcategory Konferans Öğesi - Uluslararası - Kurum Öğretim Elemanı en_US
dc.rights info:eu-repo/semantics/closedAccess en_US
dc.subject flexible substrate en_US
dc.subject intravascular ultrasonography (IVUS) en_US
dc.subject capacitive miniaturized ultrasonic transducer (CMUT) en_US
dc.title Design of capacitive micromachined ultrasonic transducers (CMUTs) on a flexible substrate for intravascular ultrasonography (IVUS) applications en_US
dc.type conferenceObject en_US